CVE-2019-25719

high
Published 2026-06-02 Β· Modified 2026-06-02
CVSS v3
8.6
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CVSS v4 NEW
8.8
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
VIR risk
8.6

Description

DrΓ€ger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Predictions

Exploit likelihood
91%
Patch ETA
β€”

Heuristic predictions, AS-IS, for prioritization only.

Mitigations

No mitigations published for this CVE yet.