CVE-2019-25723
Description
DrΓ€ger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Predictions
Heuristic predictions, AS-IS, for prioritization only.
Mitigations
No mitigations published for this CVE yet.
The vendor-content worker queues fetches as references arrive (check back in a few minutes). Or β if you've already worked around this in production β publish your fix to the community-verified tier.
β Propose a mitigation on Community β Mitigations published via the community go through AI scoring + 2 human reviewers + 7-day silent objection window before landing here withsource_tier=community-verified.
References
CWEs
CWE-1286
Community-verified mitigations for this CVE will appear above when contributors publish them.
Verify integrity in audit chain (admin only). AS-IS.