CVE-2019-25724

medium
Published 2026-06-02 Β· Modified 2026-06-02
CVSS v3
6.5
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
CVSS v4 NEW
7.1
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
VIR risk
6.5

Description

DrΓ€ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Predictions

Exploit likelihood
65%
Patch ETA
β€”

Heuristic predictions, AS-IS, for prioritization only.

Mitigations

No mitigations published for this CVE yet.

The vendor-content worker queues fetches as references arrive (check back in a few minutes). Or β€” if you've already worked around this in production β€” publish your fix to the community-verified tier.

✚ Propose a mitigation on Community β†’ Mitigations published via the community go through AI scoring + 2 human reviewers + 7-day silent objection window before landing here with source_tier=community-verified.

References

CWEs

CWE-400

Community-verified mitigations for this CVE will appear above when contributors publish them.

Verify integrity in audit chain (admin only). AS-IS.