CVE-2026-0611
Description
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Predictions
Heuristic predictions, AS-IS, for prioritization only.
Mitigations
No mitigations published for this CVE yet.
The vendor-content worker queues fetches as references arrive (check back in a few minutes). Or โ if you've already worked around this in production โ publish your fix to the community-verified tier.
โ Propose a mitigation on Community โ Mitigations published via the community go through AI scoring + 2 human reviewers + 7-day silent objection window before landing here withsource_tier=community-verified.
References
- https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel
- https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf
- https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting
CWEs
CWE-306
Community-verified mitigations for this CVE will appear above when contributors publish them.
Verify integrity in audit chain (admin only). AS-IS.