Description
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
Predictions
Exploit likelihood
97%
Patch ETA
—
Heuristic predictions, AS-IS, for prioritization only.
Mitigations
No vendor mitigations ingested yet for this CVE. The mitigation-content worker queues fetches as references arrive — check back in a few minutes, or see the references list below.
References
CWEs
CWE-259
Verify integrity in audit chain (admin only). AS-IS.